Product

Multilogger Thermometer

Multilogger Thermometer

Multilogger thermometer

Product ID: DTM-507

Temperature Datalogger DTM-507
Features:
T1, T2 for various thermocouple
Triply display with set-able backlight
Data logger function
Save Data: 128 samples with real-time data
Data Logger: 16 sets, Maximum 1024 data capacity
T/C Offset Adjust
REC, MAX, MIN, MAX-MIN, AVG, REL, HOLD functions
T1, T2, T1-T2, Time and Memory No
Resolution 0.1/1oC to 0.1/1oF
Warning beeper with Hi/Lo setting
TIME, REL, Record interval, APO time setting

Specifications:
  • DTM-507 SPECIFICATION
    • Temperature
    • Thermocouple Range
    • K-TYPE (0.1°, 1° for above 2000°F) -200°C to 1372°C,
    • -328°F to 2501°F
    • J-TYPE (0.1°, 1° for above 2000°F) -210°C to 1200°C,
    • -346°F to 2192°F
    • T-TYPE(0.1°) -200°C to 400°C, -328°F to 752°F
    • E-TYPE(0.1°) -210°C to 1000°C, -346°F to 1832°F
    • R-TYPE(1°) 0°C to 1767°C, 32°F to 3212°F
    • S-TYPE(1°) 0°C to 1767°C, 32°F to 3212°F
    • N-TYPE (0.1°, 1° for above 2000°F) -50°C to 1300°C,
    • -58°F to 2372°F
    • Accuracy: K/J/T/E-TYPE
    • ±(0.05% rdg + 0.3°C) on -50°C to 1372°C
    • ±(0.05% rdg + 0.7°C) on -50°C to -210°C
    • ±(0.05% rdg + 0.6°F) on -58°F to 2501°F
    • ±(0.05% rdg + 1.4°F) on -58°Fto -346°F
    • N-TYPE
    • ±(0.05% rdg + 0.8°C) on -50°C to 0°C
    • ±(0.05% rdg + 0.4°C) on 0°C to 1300°C
    • ±(0.05% rdg + 1.6°F) on -58°F to 32°F
    • ±(0.05% rdg + 0.8°F) on 32°F to 2372°F
    • R/S-TYPE
    • ±(0.05% rdg + 2°C) on 0°C to 1767°C
    • ±(0.05% rdg + 4°F) on 32°F to 3212°F
Features:
Q & A:

Digital thermo logger of DTM-507

Is it ok to power the meter from the 12v adaptor whilst having the 9v battery still connected, or would you recommend removing the battery?


You can use the 12V adapter, whilst having the 9V battery still connected.

This will not hurt the meter, because there is protection circuit of this temperature logger  DTM507.

If there is a break of the 12V adaptor, the meter will switch to battery power automatically

but if you use auto power off function, it will auto turn off, when the setting  time comes.

 

Contact Detail
Contact:Stanley Hwang
Address:4F.-1 , No. 225, Sec. 3, Heping E. Rd., Sinyi District, Taipei, 11056, Taiwan (R.O.C.)
TEL:886-2-27375866
FAX:886-2-27373343
Email: sales@tecpel.com
tecpelco@gmail.com
URL:http://www.oka.com.tw
http://www.tecpel.com
ZIP:11056